眶部种植体长度对骨界面应力影响的三维有限元研究

Three-dimensional finite element analysis of the effect of orbital implant lengths on stress distributions in peri-implant surfaces

作者:张兴 陈松龄 张云 齐刘英

Author:Zhang Xing, Chen Songling, Zhang Yun, Qi Liuying.

收稿日期:          年卷(期)页码:2014,32(5):484-484-487

期刊名称:华西口腔医学杂志

Journal Name:West China Journal of Stomatology

关键字:眶部种植体,应力分布,三维有限元,种植体长度,

Key words:orbital implant,stress distribution,three-dimensional finite element,implant length,

基金项目:

广东省科技厅社会发展基金资助项目(2010B060900070);国家自然科学基金面上项目资助项目(81371111)

中文摘要

目的 探讨不同长度的眶部种植体对骨界面应力分布的影响。方法 建立直径3.75 mm,长度分别为3、4、6、10 mm的眶部种植体—颅颌面骨三维有限元模型,分别给予沿种植体轴向和与轴向成45°的载荷,载荷大小20 N,记录两种方向载荷下种植体及骨界面的Von-Mises应力峰值和位移峰值,分析其应力分布。结果 施加沿种植体轴向载荷时,种植体周围应力集中于根部,种植体受力大于骨面;施加与轴向成45°载荷时,应力集中于种植体颈部与第一螺纹之间,种植体受力大于骨面。施加两个方向的载荷时,3 mm种植体的应力峰值明显大于其他长度种植体,而位移峰值无明显变化。在相同长度下,施加沿种植体轴向载荷时的应力峰值及位移峰值均明显低于与轴向成45°载荷时,载荷方式对界面应力分布有明显的影响。结论 临床上尽量选择4 mm以上的眶部种植体;应用3 mm种植体时,应选择骨密质较厚的区域植入。

英文摘要

ObjectiveThis study aims to observe the effect of orbital implant lengths on stress distribution in peri-implant surfaces.MethodsThe three-dimensional finite element analysis models of craniofacial and orbital implants with a diameter of 3.75 mm and lengths of 3, 4, 6, and 10 mm were established. A force of 20 N was applied to the models. The stress and displacement distribution under every condition were recorded and analyzed.ResultsThe loading direction along the implant axis and the stress concentration on the implant root were observed. The loading direction was at a 45 degree angle relative to the implant axis, and the stress concentration was located at the implant neck and the first screw thread. The maximum stress of the 3 mm implant was significantly higher than that under the other two loading directions. The maximum displacement of the four lengths exhibited no significant change. Given the same implant length, stress, and displacement, the peak of the implant axial direction was lower than that of the 45 degree direction. The loading type was an important factor influencing the stress and displacement of peri-implant bones.ConclusionThe implants of more than 4 mm length can be considered for clinical use. The implant of 3 mm length should be implanted in a region with thicker cortical bone.

关闭

Copyright © 2020四川大学期刊社 版权所有.

地址:成都市一环路南一段24号

邮编:610065