纯莫氏锥度连接种植系统种植体—基台锁结力及基台下沉量的实验研究

Experimental study on implant-abutment locking force and abutment subsidence in a pure Morse taper connection implant system

作者:任碧晖, 徐业豪, 戴婕婷, 郭水根, 魏洪武

Author:Ren Bihui, Xu Yehao, Dai Jieting, Guo Shuigen, Wei Hongwu

收稿日期:2023-11-17          年卷(期)页码:2024,42(3):372-372-381

期刊名称:华西口腔医学杂志

Journal Name:West China Journal of Stomatology

关键字:牙种植体,纯莫氏锥度连接,种植体—基台锁结力,基台下沉,

Key words:dental implants,pure Morse taper connection,implant-abutment locking force,abutment subsidence,

基金项目:

中文摘要

目的 探究纯莫氏锥度连接种植系统种植体与基台锁结力大小及基台下沉量的影响因素。 方法 参考Bicon种植体基台连接设计,制作不同型号的种植体试件及其对应型号的基台,种植体—基台锁结锥度统一为1.5°,锁结深度分别为1.0、2.0、3.0 mm,锁结柱直径分别为2.5、3.0、3.5 mm,种植体外壁厚度分别为0.15、0.30 mm,实验机加载力分别为200、300、400 N,每组至少10枚种植体—基台试件。所有试件均采用万能实验机进行同样的加载方式(指压+指定加载力5次),分别于指压前、指压后、实验机加压5次后测量种植体—基台总高度,计算基台下沉量,最后采用万能实验机拉开种植体与基台,观察并记录其锁结力。 结果 实验加载力、锁结深度、锁结柱直径对种植体—基台锁结力及基台下沉量均有影响,种植体—基台锁结力随实验加载力、锁结深度、锁结柱直径的增加而增加(R=0.963、R=0.607、R=0.372),其中实验加载力对种植体—基台锁结力的影响最为显著。基台下沉量随着实验加载力的增加而增加(R=0.645),随锁结深度、锁结柱直径的增加而减少(R=-0.807、R=-0.280),锁结深度对基台下沉量的影响最为明显。种植体外壁厚度与种植体—基台锁结力大小的变化无明显的相关性,但种植体外壁厚度的增加会减少基台的下沉量,两者呈反比关系。 结论 通过调整纯莫氏锥度连接种植体—基台连接设计,增加锁结深度及锁结柱直径,增加基台就位时的加载力大小及加载次数,可以增加种植体—基台锁结力,减少基台松动甚至脱落等问题。同时为防止基台下沉导致后期咬合关系的改变,建议基台就位时加载次数不少于5次,同时建议使用临时修复体及早期只行初步咬合调整,在使用一段时间后再行最终修复及最终的咬合调整。

英文摘要

ObjectiveThis test aimed to investigate the factors affecting the locking force between the implant and abutment and the amount of abutment subsidence in pure Morse taper connection implant systems.MethodsWith reference to the Bicon implant abutment connection design, different types of implant specimens and their corresponding types of abutments were fabricated. The implant-abutment locking taper was uniformly 1.5°. The locking depths were 1.0, 2.0, and 3.0 mm. The diameters of the locking column were 2.5, 3.0, and 3.5 mm. The thicknesses of the outer wall of the implant were 0.15 and 0.30 mm. The loading forces of the testing machine were 200, 300, and 400 N. At least 10 specimens of each group of implant-abutment were used. All specimens were loaded in the same manner using a universal testing machine (finger pressure + specified loading force, five times). The total height of the implant-abutment was measured before finger pressure, after finger pressure, and after the testing machine was loaded for five times to calculate the amount of sinking of the abutment. Finally, the implant and abutment were pulled apart using the universal testing machine, and the subluxation force was observed and recorded.ResultsThe test loading force, locking depth, and locking post diameter had an effect on the implant-abutment locking force and abutment subsidence. The implant-abutment locking force increased with the increase in the test loading force, locking depth, and locking post diameter (R=0.963, 0.607, and 0.372, respectively), with the test loading force having the most significant effect. Abutment subsidence increased with the increase in test loading force (R=0.645) and decreased with the increase in locking depth and locking post diameter (R=-0.807 and -0.280, respectively), with locking depth having the most significant effect on abutment subsidence. No significant correlation was found between the thickness of the outer wall of the implant and the change in the magnitude of the implant-abutment locking force. However, an increase in the thickness of the outer wall of the implant decreased the amount of abutment subsidence, which was inversely correlated.ConclusionThe locking force of the implant-abutment can be increased by adjusting the design of the pure Morse taper connection implant⁃abutment connection, increasing the locking depth and locking post diameter, and increasing the amount and number of times the abutment is loaded during seating. Problems, such as loosening or detachment of the abutment, can be reduced. The recommended abutment to be loaded should be no less than five times during seating to prevent the abutment from sinking and causing changes in the occlusal relationship in the later stages. Preliminary occlusal adjustments should only be conducted in the early stages of the use of temporary restorations, and final restorations and occlusal adjustments are recommended to be performed after using the abutment for a period of time.

上一条:一种通用型牙种植导板在牙模上模拟后牙区种植的精度评价

关闭

Copyright © 2020四川大学期刊社 版权所有.

地址:成都市一环路南一段24号

邮编:610065