期刊导航

论文摘要

3D打印树脂种植工作模型尺寸稳定性对替代体复制精度影响的初步研究

Preliminary study on the influence of the dimensional stability of 3D printed resin master model on the replication accuracy of implant replicas

作者:李欣, 鲁玉宗, 杨咏涛, 温奥楠, 王勇, 赵一姣

Author:Li Xin, Lu Yuzong, Yang Yongtao, Wen Aonan, Wang Yong, Zhao Yijiao

收稿日期:2024-12-18          年卷(期)页码:2025,43(5):689-689-695

期刊名称:华西口腔医学杂志

Journal Name:West China Journal of Stomatology

关键字:3D打印,树脂工作模型,尺寸稳定性,种植修复,精度,

Key words:3D printing,resin working model,dimensional stability,implant restoration,accuracy,

基金项目:青岛市自然科学基金(23-2-133-zyyd-jch);山东省医药卫生科技发展计划项目(202208050566)

中文摘要

目的 研究3D打印树脂种植工作模型尺寸稳定性对替代体复制精度的影响。 方法 获取10例拟进行2个种植体支持的联冠或者固定桥修复患者的数字化印模,在模型打印完成后的3、7、14 d在替代体上安装扫描杆进行扫描,在Geomagic Wrap 2021软件中基于扫描数据分析不同时间扫描杆相对于口内扫描的3D偏差、线性偏差和角度偏差。 结果 种植替代体位置发生了与模型收缩方向一致的近中舌侧偏移。从第7天开始,替代体(扫描杆)的3D偏差、距离线性偏差和角度偏差相较于第3天显著增大(P<0.05)。在第14天时,变化更加明显,其中3D偏差、距离线性偏差及轴线角度偏差与3 d和7 d组相比差异均具有统计学意义(P<0.05)。各组间高度线性偏差差异无统计学意义。 结论 3D打印树脂模型的尺寸稳定性不足会导致替代体相对位置和角度的变化,从而影响替代体对种植体位置复制的精度。建议在模型打印完的7 d内完成修复体的制作。

英文摘要

ObjectiveThis study aimed to investigate the influence of the dimensional stability of 3D printed resin master model on the replication accuracy of implant replicas.MethodsTen digital impressions of patients undergoing continuous crowns or fixed bridge restoration supported by two implants were obtained, and resin models with implant replicas were 3D printed. Scanning rods were fixed on the replicas 3, 7, and 14 days after printing. The 3D, linear, and angular deviations of the scanning rods at different times were analyzed through Geomagic Wrap 2021 software.ResultsThe position of the replicas shifted mesiolingually, in the same direction as the shrinkage of the model. From day 7 onward, the 3D, distance linear, and angular deviations of the replicas (scanning rod) significantly increased compared with those on the 3rd day (P<0 .05). on the 14th day, the changes were even more pronounced, with the above deviations showing statistical significance (P<0 .05) compared with those for the 3-day and 7-day groups. no statistical difference in height linear deviation was observed among the groups.ConclusionThe insufficient dimensional stability of 3D printed resin models can lead to changes in the relative position and angle of the replicas, thereby affecting the accuracy of the replicas in recreating the implant’s position. Complete manufacturing of prosthesis is recommended within 7 days after the model is printed.

上一条:梯形瓣和角形瓣对引导骨再生术后黏膜血供及成骨效果的影响

下一条:墨旱莲—女贞子治疗牙周炎的潜在成分及作用机制

关闭

Copyright © 2020四川大学期刊社 版权所有.

地址:成都市一环路南一段24号

邮编:610065